== Estimating Input Delay and Rate Constants == | == Estimating Input Delay and Rate Constants == | ||
This example demonstrates an approach to simultaneously estimating input function delay along with parameters of the 1-tissue compartment (e.g. blood flow) model. | This example demonstrates an approach to simultaneously estimating input function delay along with parameters of the 1-tissue compartment (e.g. blood flow) model. | ||
− | For the sake of generality, this could be interpreted as a | + | For the sake of generality, this could be interpreted as a dynamic contrast-enhanced (DCE) MRI or perfusion PET study. |
Line 15: | Line 15: | ||
− | where C<sub>T</sub> is the tissue concentration and C<sub>p</sub> is the plasma concentration | + | where C<sub>T</sub> is the tissue concentration and C<sub>p</sub> is the plasma concentration. |
C<sub>T</sub> and C<sub>p</sub> are interpreted as either molar concentrations ([http://www.ncbi.nlm.nih.gov/pubmed/17555251?ordinalpos=2&itool=EntrezSystem2.PEntrez.Pubmed.Pubmed_ResultsPanel.Pubmed_RVDocSum Salinas, Muzic and Saidel 2007]). | C<sub>T</sub> and C<sub>p</sub> are interpreted as either molar concentrations ([http://www.ncbi.nlm.nih.gov/pubmed/17555251?ordinalpos=2&itool=EntrezSystem2.PEntrez.Pubmed.Pubmed_ResultsPanel.Pubmed_RVDocSum Salinas, Muzic and Saidel 2007]). | ||
Line 22: | Line 22: | ||
This example also demonstrates how to temporally align measured input function data to the tissue (image) data. Here the plasma concentration vs. time t is modeled as | This example also demonstrates how to temporally align measured input function data to the tissue (image) data. Here the plasma concentration vs. time t is modeled as | ||
− | C<sub>p</sub>(t, tau) = M(t- | + | C<sub>p</sub>(t, tau) = M(t-tau) if t >= tau; 0 if t < tau . |
M(t) is the measured input data, tau is the delay parameter to estimate, and t is the time in minutes. | M(t) is the measured input data, tau is the delay parameter to estimate, and t is the time in minutes. | ||
Line 28: | Line 28: | ||
Because the delay tau is estimated in this example, it will be handy to note that derivative of the input with respect to delay is | Because the delay tau is estimated in this example, it will be handy to note that derivative of the input with respect to delay is | ||
− | + | dC<sub>p</sub>(t)/d tau = - dM/dt|<sub>t-tau</sub> | |
For this example, M is determined by using linear interpolation to interpolate between measured values. The interpolation is implemented using piecewise polynomial framework (and it could be easily modified to accommodate cubic spline interpolation). | For this example, M is determined by using linear interpolation to interpolate between measured values. The interpolation is implemented using piecewise polynomial framework (and it could be easily modified to accommodate cubic spline interpolation). | ||
Line 66: | Line 66: | ||
end | end | ||
</pre> | </pre> | ||
− | + | ||
− | ==== | + | == Example: Estimating Input Delay and Rate Constants == |
− | In this example, the | + | |
+ | In this example, the model output are assumed to represent measurements of C<sub>T</sub>. (To keep the example simple, intravascular concentration of tracer is ignored.) | ||
The values of the parameters p = [K<sub>1</sub>; k<sub>2</sub>; tau] will be estimated. | The values of the parameters p = [K<sub>1</sub>; k<sub>2</sub>; tau] will be estimated. | ||
+ | '''Step 1''' To set the simulation in Matlab, create the above function and put in in a file called DelayExampleInput.m. | ||
− | + | '''Step 2''' Create a 1-tissue compartment model | |
<pre> | <pre> | ||
% create new, empty model | % create new, empty model | ||
Line 85: | Line 87: | ||
cm = addParameter(cm, 'dk', 0.34); % decay constant | cm = addParameter(cm, 'dk', 0.34); % decay constant | ||
cm = addParameter(cm, 'tau', 0.25); % time delay | cm = addParameter(cm, 'tau', 0.25); % time delay | ||
+ | |||
% add compartments | % add compartments | ||
cm = addCompartment(cm, 'CT'); | cm = addCompartment(cm, 'CT'); | ||
Line 91: | Line 94: | ||
% define the input(first column is time, second it concentration) | % define the input(first column is time, second it concentration) | ||
inputData = [ | inputData = [ | ||
− | 0 | + | 0 0 |
0.0500 0 | 0.0500 0 | ||
0.1000 0 | 0.1000 0 | ||
Line 111: | Line 114: | ||
% determine spline coefficients for linear interpolation | % determine spline coefficients for linear interpolation | ||
ppM = lspline(inputData(:,1), inputData(:,2)); | ppM = lspline(inputData(:,1), inputData(:,2)); | ||
+ | |||
% analytically calculate derivative | % analytically calculate derivative | ||
[breaks, coefs, l, k, d] = unmkpp(ppM); | [breaks, coefs, l, k, d] = unmkpp(ppM); | ||
Line 128: | Line 132: | ||
st = [[0:5:85]' [5:5:90]']/60; % division by 60 converts sec to min | st = [[0:5:85]' [5:5:90]']/60; % division by 60 converts sec to min | ||
cm = set(cm, 'ScanTime', st); | cm = set(cm, 'ScanTime', st); | ||
+ | midScanTime = (st(:,1) + st(:,2)) / 2; | ||
− | % solve the model | + | % solve the model and generate the output |
[sol,solIndex] = solve(cm); | [sol,solIndex] = solve(cm); | ||
+ | plot(midScanTime,sol(:,3),'LineWidth',2); | ||
+ | xlabel('Minutes'); | ||
+ | ylabel('KBq/ml'); | ||
+ | </pre> | ||
+ | |||
+ | [[Image:Pretest.jpeg]] | ||
− | + | The above figure is the model output (C<sub>T</sub>). Here, we use it as perfect experimental data that would be fitted. | |
− | % | + | '''Step 3''' Use model output as measured C<sub>T</sub> (perfect "data") and fit it |
− | pinit=[0. | + | <pre> |
+ | % Set parameters: initial guess, lower and upper bounds for K1, k2 and tau | ||
+ | pinit=[0.1;0.2;0.1]; | ||
lb =[0.01;0.01;0.01]; | lb =[0.01;0.01;0.01]; | ||
ub =[1;1;5]; | ub =[1;1;5]; | ||
+ | % specify parameters to be adjusted in fitting | ||
cm=addSensitivity(cm,'k1','k2','tau'); | cm=addSensitivity(cm,'k1','k2','tau'); | ||
+ | % Experimental Data to be fit | ||
cm=set(cm,'ExperimentalData',sol(:,3)); | cm=set(cm,'ExperimentalData',sol(:,3)); | ||
− | |||
− | % | + | % Perform curve fitting |
+ | [pfit,qfitnull,modelfit,pp1,output]=fitGen(cm,pinit,lb,ub,'OLS'); | ||
+ | </pre> | ||
+ | |||
+ | '''Step 4''' Examine model outputs using estimated parameter (pfit) and initial guess (pinit) | ||
+ | <pre> | ||
+ | % Solve model using estimated parameters | ||
cm1=cm; | cm1=cm; | ||
− | cm1=set(cm1,'ParameterValue','k1', | + | cm1=set(cm1,'ParameterValue','k1',pfit(1)); |
− | cm1=set(cm1,'ParameterValue','k2', | + | cm1=set(cm1,'ParameterValue','k2',pfit(2)); |
− | cm1=set(cm1,'ParameterValue','tau', | + | cm1=set(cm1,'ParameterValue','tau',pfit(3)); |
est_sol=solve(cm1); | est_sol=solve(cm1); | ||
− | plot(midScanTime,sol(:,3),'o',midScanTime,est_sol(:,3),'r-'); | + | % solve model using initial parameters |
+ | cm1=set(cm1,'ParameterValue','k1',pinit(1)); | ||
+ | cm1=set(cm1,'ParameterValue','k2',pinit(2)); | ||
+ | cm1=set(cm1,'ParameterValue','tau',pinit(3)); | ||
+ | est_sol_init=solve(cm1); | ||
+ | |||
+ | plot(midScanTime,sol(:,3),'o',midScanTime,est_sol(:,3),'r-',midScanTime,est_sol_init(:,3),'g-','LineWidth',2); | ||
xlabel('Minutes'); | xlabel('Minutes'); | ||
ylabel('KBq/ml'); | ylabel('KBq/ml'); | ||
− | legend('Data', 'Fit'); | + | legend('Data', 'Fit','Initial Guess'); |
</pre> | </pre> | ||
− | [[Image: | + | [[Image:test.jpeg]] |
− | |||
− |
Latest revision as of 03:46, 23 December 2008
Estimating Input Delay and Rate Constants
This example demonstrates an approach to simultaneously estimating input function delay along with parameters of the 1-tissue compartment (e.g. blood flow) model. For the sake of generality, this could be interpreted as a dynamic contrast-enhanced (DCE) MRI or perfusion PET study.
Tissue uptake
This model has one tissue compartment. Material in the blood is assumed to rapidly exchange with that in (extravascular) tissue. The tissue model has two rate constants: K1 and k2 and is depicted in the diagram:
It can also be described by the differential equation:
dCT/dt = K1 Cp - k2 CT
where CT is the tissue concentration and Cp is the plasma concentration.
CT and Cp are interpreted as either molar concentrations (Salinas, Muzic and Saidel 2007).
Input function
This example also demonstrates how to temporally align measured input function data to the tissue (image) data. Here the plasma concentration vs. time t is modeled as
Cp(t, tau) = M(t-tau) if t >= tau; 0 if t < tau .
M(t) is the measured input data, tau is the delay parameter to estimate, and t is the time in minutes.
Because the delay tau is estimated in this example, it will be handy to note that derivative of the input with respect to delay is
dCp(t)/d tau = - dM/dt|t-tau
For this example, M is determined by using linear interpolation to interpolate between measured values. The interpolation is implemented using piecewise polynomial framework (and it could be easily modified to accommodate cubic spline interpolation).
Let inputData be a two-column matrix with column 1 holding sample times and column 2 holding the sampled input concentrations.
The interpolation coefficients may be calculated using the lspline function
ppM = lspline(inputData(:,1), inputData(:,2));
To evaluate the derivative dCp/dt, which can be determined in terms of dM/dt by analytically differentiating the interpolating polynomial.
[breaks, coefs, l, k, d] = unmkpp(ppM); ppdMdtau = mkpp(breaks, repmat(k-1:-1:1,d*l,1) .* coefs(:,1:k-1), d);
The input function is then implemented as
function [Cp, dCpdtau] = DelayExampleInput(parm, t, X) if (nargout > 0), t = t(:); % ensure t is a column vector tau = parm(1); % delay ppC = X{1}; % piecewise-polynomial coefficients for Cp Cp = zeros(size(t)); f = find(t > tau); Cp(f) = ppval(ppC, t(f) - tau); if (nargout > 1), ppdCdt = X{2}; % piecewise-polynomial coefficients for derivative dCpdtau = zeros(size(t)); dCpdtau(f) = -ppval(ppdCdt, t(f) - tau); end end
Example: Estimating Input Delay and Rate Constants
In this example, the model output are assumed to represent measurements of CT. (To keep the example simple, intravascular concentration of tracer is ignored.) The values of the parameters p = [K1; k2; tau] will be estimated.
Step 1 To set the simulation in Matlab, create the above function and put in in a file called DelayExampleInput.m.
Step 2 Create a 1-tissue compartment model
% create new, empty model cm = compartmentModel; % define default values for parameters cm = addParameter(cm, 'k1', 0.3); cm = addParameter(cm, 'k2', 0.5); cm = addParameter(cm, 'tau', 0.25); cm = addParameter(cm, 'sa', 1); % specific activity at t=0 cm = addParameter(cm, 'dk', 0.34); % decay constant cm = addParameter(cm, 'tau', 0.25); % time delay % add compartments cm = addCompartment(cm, 'CT'); cm = addCompartment(cm, 'J'); % define the input(first column is time, second it concentration) inputData = [ 0 0 0.0500 0 0.1000 0 0.1500 14.5234 0.2000 52.1622 0.2500 76.6730 0.3000 91.6416 0.4000 103.0927 0.5000 100.9401 0.7000 83.5343 0.8000 74.3628 1.0000 59.7726 1.2500 48.7530 1.5000 43.0772 1.7500 40.1924 2.0000 38.6236 ]; % determine spline coefficients for linear interpolation ppM = lspline(inputData(:,1), inputData(:,2)); % analytically calculate derivative [breaks, coefs, l, k, d] = unmkpp(ppM); ppdMdtau = mkpp(breaks, repmat(k-1:-1:1,d*l,1) .* coefs(:,1:k-1), d); X = {ppM, ppdMdtau}; cm = addInput(cm, 'Cp','sa','dk', 'DelayExampleInput', 'tau', X); % connect compartments and inputs cm = addLink(cm, 'L', 'Cp', 'CT', 'k1'); cm = addLink(cm, 'K', 'CT', 'J', 'k2'); % define the activity concentration in tissue pixel cm = addOutput(cm, 'TissuePixel', {'CT','1'}, {'Cp','0'}); % specify scan frame times st = [[0:5:85]' [5:5:90]']/60; % division by 60 converts sec to min cm = set(cm, 'ScanTime', st); midScanTime = (st(:,1) + st(:,2)) / 2; % solve the model and generate the output [sol,solIndex] = solve(cm); plot(midScanTime,sol(:,3),'LineWidth',2); xlabel('Minutes'); ylabel('KBq/ml');
The above figure is the model output (CT). Here, we use it as perfect experimental data that would be fitted.
Step 3 Use model output as measured CT (perfect "data") and fit it
% Set parameters: initial guess, lower and upper bounds for K1, k2 and tau pinit=[0.1;0.2;0.1]; lb =[0.01;0.01;0.01]; ub =[1;1;5]; % specify parameters to be adjusted in fitting cm=addSensitivity(cm,'k1','k2','tau'); % Experimental Data to be fit cm=set(cm,'ExperimentalData',sol(:,3)); % Perform curve fitting [pfit,qfitnull,modelfit,pp1,output]=fitGen(cm,pinit,lb,ub,'OLS');
Step 4 Examine model outputs using estimated parameter (pfit) and initial guess (pinit)
% Solve model using estimated parameters cm1=cm; cm1=set(cm1,'ParameterValue','k1',pfit(1)); cm1=set(cm1,'ParameterValue','k2',pfit(2)); cm1=set(cm1,'ParameterValue','tau',pfit(3)); est_sol=solve(cm1); % solve model using initial parameters cm1=set(cm1,'ParameterValue','k1',pinit(1)); cm1=set(cm1,'ParameterValue','k2',pinit(2)); cm1=set(cm1,'ParameterValue','tau',pinit(3)); est_sol_init=solve(cm1); plot(midScanTime,sol(:,3),'o',midScanTime,est_sol(:,3),'r-',midScanTime,est_sol_init(:,3),'g-','LineWidth',2); xlabel('Minutes'); ylabel('KBq/ml'); legend('Data', 'Fit','Initial Guess');